Journal Archive

Johnson Matthey Technol. Rev., 2023, 67, (1), 4
doi: 10.1595/205651323X16601466421853

Step-Change Improvements in Traceable Process Control Thermometry

Key to product quality, safety and sustainability


  • Jonathan V. Pearce*, Declan L. Tucker, Radka I. Veltcheva, Graham Machin
  • National Physical Laboratory, Hampton Road, Teddington, TW11 0LW, UK
  • *Email: jonathan.pearce@npl.co.uk

PEER REVIEWED 
Received 1st June 2022; Revised 8th August 2022; Accepted 10th August 2022; Online 11th August 2022


Article Synopsis

Measurement and control of process temperature is key to maximising product quality, optimising efficiency, reducing waste, safety and minimising carbon dioxide and other harmful emissions. Drift of temperature sensor calibration due to environmental factors such as high temperature, vibration, contamination and ionising radiation results in a progressively worsening temperature measurement error, which in turn results in suboptimal processes. Here we outline some new developments to overcome sensor calibration drift and so provide assured temperature measurement in process, including self-validating thermocouples, embedded temperature reference standards, and practical primary Johnson noise thermometry where the temperature is measured directly without the need for any calibration. These new developments will give measurement assurance by either providing measurements which are inherently stable, or by providing an in situ calibration facility to enable the detection and correction of calibration drift.

Introduction

Temperature is extremely difficult to measure directly, so most practical thermometry is performed with devices that have some temperature-dependent property (such as resistance or voltage). Such devices must be calibrated to establish the relationship between this property and the temperature.

The global temperature measurement infrastructure is essential for ensuring equivalence between temperature measurements worldwide. It is currently evolving in a way that will open up new opportunities for improved temperature measurement in process, through the move from a definition based on a physical artefact (the water triple point cell) to one based on a fundamental constant (1, 2). The current framework used for approximating temperature is provided by the International Temperature Scale of 1990 (ITS-90) (3). Specifically, the ITS-90 provides the framework for tracing temperature measurements back to the International System of Units (SI) unit of temperature, the kelvin, in a practical way. When calibration activities are accredited by a national accreditation body to ISO 17025:2017 (4) this provides assurance that the measurements are reliable and traceable to the SI. There are two key components (5).

The first component is a set of specific temperature values, called fixed points, based on highly reproducible phase changes of materials. These temperatures have been assigned a priori by international consortia using very accurate thermodynamic thermometry techniques to measure their temperature directly. The second component is a set of specified interpolating devices and associated equations which provide low interpolation uncertainties at temperatures between the fixed points.

A key drawback of this approach is that harsh environments can physically change the temperature sensor and hence the relationship between the thermometer output and temperature, which in turn causes a loss of calibration and therefore measurement traceability. In this article some solutions to this problem are outlined whereby traceability can be maintained in process.

This is linked to the ‘reproducibility crisis’ (6, 7) whereby two temperature measurements of the same phenomenon taken by two different groups differ, often by unacceptably large amounts. It is often impossible to know which one is more trustworthy, often because no clear record of measurement traceability is available.

In this article the problem of calibration drift is described, then some practical solutions to retain traceability in process are described, namely self-validating thermocouples, embedded temperature references and finally a paradigm shift in the form of practical ‘primary’ thermometry whereby temperature is measured directly.

Calibration Drift

When temperature sensors are used to monitor processes in harsh environments they can be subjected to contamination (for example in a jet engine or in a silicon processing plant), high temperatures (such as during casting of single crystal nickel alloy turbine blades or quartz manufacturing), vibration (during flight or launch of a spacecraft) or ionising radiation (mainly nuclear applications). The result is that the composition or microstructure of the sensor changes. This results in a change in the relationship between the sensor output (such as voltage or resistance) and the temperature, which is referred to as calibration drift. This in turn causes a progressive loss of knowledge of the process temperature by an unknown amount, resulting in a measurement error and an (often unnoticed) change in the process parameters.

It cannot be emphasised enough that in practice there is no indication of this calibration drift occurring. It causes inefficiency in terms of higher than necessary energy consumption, poor yield or unnecessarily large product rejection and waste, and in some cases inability to achieve the desired regions of the process envelope. It can also have safety implications necessitating suboptimal operation, for example in processes involving combustion, steam generation or other high temperature aspects which become more efficient as the temperature increases, but which are run at conservatively low temperatures to preserve a large safety margin. In many process control applications significant resources are spent on replacing sensors at considerable cost, in terms of expense and process down time.

Self-Validating Thermocouples

Thermocouples are very widely used in industry for process monitoring and control. They are particularly susceptible to calibration drift, because they tend to be used in environments which are harsh. This can be overcome by equipping the thermocouple with a miniature fixed point, which comprises of a metal (in a graphite container) with a known melting temperature. The fixed point is placed in contact with the thermocouple measurement junction, so that when the ambient temperature exceeds the melting temperature of the fixed point, a ‘hesitation’ is seen in the thermocouple output corresponding to the melting plateau arising from the absorption of heat from the surroundings during melting. Since the temperature during this hesitation is known a priori, the thermocouple can be recalibrated in situ, allowing the calibration drift to be determined.

The choice of phase-change material will depend on the operating temperature envisaged; it can be selected from the ITS-90 defining fixed point metals (3), or, for temperatures above about 1100°C, high temperature fixed points based on metal-carbon alloys can be used (814). These materials are summarised in Table I.

Table I

Fixed Point Materials with their Melting Temperaturesa

Fixed point Melting temperature, °C
Pure metals
Ga 29.7646
In 156.5985
Sn 231.928
Zn 419.527
Al 660.323
Ag 961.78
Au 1064.18
Cu 1084.62
Metal-carbon alloys
Fe-C 1153
Co-C 1324
Ni-C 1329
Pd-C 1492
Rh-C 1657
Pt-C 1738
Ru-C 1954
Ir-C 2291
Re-C 2474
WC-C 2749

aDefinitive values for the melting temperatures of Co-C, Pt-C and Re-C values are given in (10). Low uncertainty values for Fe-C, Pd-C, Ru-C and the WC-C point are being evaluated as part of the EMPIR Real-K project

A number of approaches to self-validating thermometers have been made. One of the first well-characterised self-validating thermocouples was developed by Tischler et al. (15) and other effective ones were subsequently demonstrated (1619), including one which was trialled in a power generation application (20). The concept was extended to higher temperatures using high temperature metal-carbon fixed points (21) and ultimately to 2300°C (22). A phase-change cell based on the Curie point (the temperature above which some materials lose their permanent magnetic properties, associated with a detectable hesitation in the temperature) was recently incorporated with a resistance thermometer to provide assurance in hygiene critical applications; this is now commercially available (23).

The challenge is to make these devices small enough to fit inside typical protective sheaths used in industrial thermocouples, so that the self-validating thermocouples are, from a practical point of view, indistinguishable from regular process control thermocouples in terms of incorporation in the process environment.

A practical self-validating thermocouple concept has been developed by the National Physical Laboratory (NPL), UK, and is now being commercialised by UK thermocouple manufacturer CCPI Europe Ltd (24, 25) with the trade name INSEVA. Such a device is shown in Figure 1. The long-term uncertainty achievable with the self-validating thermocouple is approximately 0.5°C (25). It is currently being trialled in heat treatment applications where the process temperature must remain within a very tight process control envelope. Some example melting curves obtained with the INSEVA device (using the silver melting point, 961.78°C and the gold melting point, 1064.18°C) are shown in Figure 2.

Fig. 1.

(a) Schematic of self-validating thermocouple, which comprises a regular noble metal thermocouple with a small fixed-point cell positioned at the tip of the thermocouple (image courtesy of CCPI Europe Ltd, UK); (b) a £2 coin is shown for scale (which is of a similar size to a €2 coin)

Fig. 2.

Melting curves yielded by the INSEVA thermocouple in various processes, using: (a)–(c) the silver melting point (961.78°C); and (d) gold melting point (1064.18°C). The horizontal lines show the melting point and the upper and lower limits of the uncertainty (approximately 95% coverage probability) in the determination thereof. Note in (b) the ‘glitch’ at about 750°C is ignored by the prototype automatic plateau detection algorithm, indicating its robustness. The plot in (c) shows the thermocouple indication in terms of its thermovoltage rather than temperature, but the same principles apply

Current development activity is focused on proving the technology in industrial trials, so far in heat treatment applications in the UK and Europe, and on the development of algorithms for real-time automatic detection and characterisation of the melting plateau to facilitate recalibration in software. Early results with trials at process temperatures up to around 1200°C with the silver, gold and copper fixed points are promising, with the NPL algorithm reliably detecting and characterising a wide range of melting plateaus of different shapes; these all rely on characterising the density of points to produce a histogram in terms of temperature which can then be characterised to determine the melting point (26). A hybrid approach is taken due to the wide variation of the shape of melting plateaus (Figure 2) depending on the ambient temperature conditions (for example, the shape of the melting curve will vary depending on whether the ambient temperature is subjected to constant ramp rate, or held constant, or other condition), so that in some cases the histogram approach is combined with a correctional algorithm based on detection of points rising above the noise (27). Further development is currently underway, with efforts focusing on machine learning for fully autonomous detection and characterisation of the melting plateau.

Embedded Reference Phase-Change Cells

A similar concept has been developed for instrumentation used on board spacecraft (28, 29). The concept of using small phase-change cells for SI traceable calibrations on board spacecraft was discussed by Sapritsky (30). Devices based on gallium and gallium-based alloys have been developed by Sun et al. and Hao et al. (31, 32) with a length and diameter of 25 mm and 80 mm respectively. Similar size (containing about 100 g of phase-change material) gallium and gallium alloy cells for space applications have been developed by Burdakin et al. (33, 34) and by Gero et al. (35) for both contact and non-contact thermometry which claimed a realisation uncertainty of about 5 mK. All these cells are still too large for the RAL Space (Science and Technology Facilities Council, UK) application. Topham et al. developed feasible miniature references (36) containing about 4 g of gallium and stainless steel bellows to accommodate expansion and contraction of the phase-change material, and showed during tests on the International Space Station that the melting and freezing of gallium is not affected by weightless conditions (37). During those tests the device worked well although a number of potential improvements were identified.

NPL and RAL Space have collaborated on the development of a robust miniature phase-change cell which is robust and can operate autonomously. This makes use of the melting point of gallium, 29.7646°C, which is a convenient reference temperature for applications in Earth observations. The miniature phase-change cell contains about 2 g of gallium. The cell, which is made of stainless steel (Figure 3), is embedded in a prototype calibration blackbody structure adjacent to the platinum resistance thermometers used for measuring the temperature of the blackbody. The blackbody is used for calibrating radiometric instrumentation for climate monitoring, so errors arising from calibration drift have serious implications for these measurements, which feed into datasets used for high profile applications such as environmental policy development.

Fig. 3.

(a) Miniature phase-change cell containing gallium (melting point 29.7646°C); (b) computer-aided design (CAD) drawing of the miniature gallium phase-change cell (blue) embedded in the calibration blackbody. The platinum resistance thermometer to be calibrated is located to the left of the cell

Key technical challenges in developing a practical device were: (a) to prevent a reaction between the gallium and the steel walls of the container (which could become a problem over the very long shelf-life of many years required for this instrumentation); and (b) to ensure reliable melting and freezing of the gallium (which has a pronounced tendency to remain liquid at temperatures many degrees below its freezing temperature). Another key part of the container is the weld which is essential to hermetically seal the cell and prevent any leakage of gallium, which could have a detrimental effect on the aluminium components of the spacecraft.

The cell yields very clearly defined melting curves (Figure 4), which last several hours, and which have a melting range of less than 0.01°C. This is remarkable performance given the tiny size of the phase-change cell, and that the thermometers are displaced from it rather than immersed in it (immersion of the thermometer in the phase-change cell being normal for general high precision temperature metrology applications), although it should be noted that the apparent (i.e. indicated) plateau temperature will have some dependence on the ambient temperature and its rate of change; this can be mitigated by characterising this dependence a priori, or by judicious selection of the part of the melting curve; the earlier stages of melting are the least susceptible to this influence, as can be seen in Figure 2. By calibrating the cell against NPL’s standard reference gallium cell, it can be used to calibrate the resistance thermometers on board the spacecraft with uncertainty of less than 0.01°C. Current work is focused on the development of additional fixed points between 0°C and 30°C.

Fig. 4.

Several melting plateaus for the miniature gallium phase-change cell, as indicated by the adjacent platinum resistance thermometer, illustrating the repeatability

Although this technique was developed for space applications, it is more widely applicable, for example for subsea temperature measurements where accuracy is of crucial importance. Deep ocean temperature measurements are extremely difficult and are heavily reliant on good reproducibility, as they need to be able to detect changes in ocean temperature over decades which amount to a few thousandths of a degree (38). This is very important because such a small temperature change of the ocean represents a vast change of the stored energy. Fixed points at the triple points of water (0.01°C) and mercury (–38.8344°C) would be useful for increasing confidence that the very small observed temperature changes reported are in fact real and not due to small calibration shifts of the sensors.

Practical Johnson Noise Thermometry

Another way to overcome the problem of calibration drift is to dispense with the need for calibration altogether by measuring temperature directly. This is referred to as practical primary thermometry, and the redefinition of the SI unit of temperature, the kelvin, in 2019 enables the kelvin to be realised in situ rather than relying on a traceability chain (1, 39).

If a property can be linked to temperature through fundamental physics, without recourse to a temperature scale or calibration, and if all parameters needed to work out the temperature are continuously measured, then a device can be used to measure this property which does not suffer from calibration drift, because any change or degradation of the sensor material is accounted for in the measurement. One such parameter is Johnson noise, which is the extremely small voltage arising from the random thermal motion of electrons in a resistor. A comprehensive review of Johnson noise thermometry is given in (40). The fact that the thermal motion of the electrons becomes more vigorous with increasing temperature results in the linkage between Johnson noise voltage and temperature via Nyquist’s equation, Equation (i) (41):

(i)

where is the mean squared Johnson noise voltage, k is the Boltzmann constant, R is the sensor resistance and Δf is the frequency bandwidth which depends on the properties of the sensing electronics and cables. A crucial point is that if the resistance R is measured continuously then all properties of the sensor and cabling which are affected by the harsh environment are characterised. This means that the thermometer can be used to measure temperature reliably at all times, independently of any calibration and, importantly, regardless of any sensor and cable deterioration.

The measurement of the Johnson noise voltage, which is miniscule (typically of the order of 1 μV root mean square even with a wide bandwidth of 1 MHz) is very challenging, especially in the presence of far larger external noise sources and has until recently been the preserve of large national laboratories concerned with measurement standards activities (42). Considerable design effort has been devoted to keeping external noise out of the system (43).

There have been many attempts to develop a practical Johnson noise thermometer. Brixy et al. (44) developed a thermometer which comprised of a thermocouple for rapid response and a Johnson noise thermometer for long-term stability, with the four wires used for Johnson noise thermometer consisting of two thermocouples. This was successfully tested in nuclear environments, hot isostatic pressing, glass manufacturing and petrochemical applications, operating up to 2200°C (45). Other types of dual probes were developed by Soulen et al. for cryogenic applications (46), Blalock et al. for nuclear applications at 300°C (47), de Groot et al. for high temperature applications up to 1500°C (48) (although stability was a problem in the latter application) and Borkowski et al. (49). Extensive work has been performed on Johnson noise thermometer for nuclear applications at Forschungszentrum Jülich in Germany (45, 50) although this is now discontinued, and at Oak Ridge National Laboratory in the USA (see for example (51)). Most recently, a new approach has been applied to develop a practical high accuracy Johnson noise thermometer using a range of completely new techniques to overcome lead wire effects and amplifier noise (52).

However, so far no practical Johnson noise thermometer has reached the market. A current collaboration between Metrosol Ltd (a UK instrumentation electronics specialist) and NPL has overcome this challenge and developed a prototype thermometer which employs completely new approaches using analysis in the frequency domain rather than the more conventional time domain. This enables simultaneous measurement of the resistance of the sensing element and the thermal noise power, as well as calibration of the noise power measurement and enabling a wider bandwidth (and hence higher signal to noise ratio) by compensating for the frequency dependent effect of the cables.

The device has unprecedented immunity to external electrical interference (53, 54), passing the most stringent heavy industrial immunity standard test required by IEC 61000-4-3:2020 (55). The accuracy achieved is about ±0.5°C in a few seconds (54). The current prototype is shown in Figure 5.

Fig. 5.

(a) Prototype practical Johnson noise thermometer developed by Metrosol in collaboration with NPL, showing the probe and sensor head containing the electronics (image courtesy of Metrosol Ltd, UK); (b) standard deviation of temperature measurements of the prototype practical Johnson noise thermometer compared with the theoretical ‘noise floor’ for the same averaging time (which in this case is 7 s)

Figure 5 also shows the standard deviation of the temperature measured with the practical Johnson noise thermometer prototype up to 1000°C, compared with the limit given by Rice’s equation (56, 57), which represents the noise ‘floor’ below which the standard deviation of no measurement can go. It can be seen that the measurements are already quite close to the noise floor; the principal way of improving the statistical quality (and hence the measurement uncertainty) further is by averaging the measurement over a longer duration, so Johnson noise thermometer measurements always represent a compromise between accuracy and response time.

For Johnson noise thermometry, the traceability is transferred from temperature to electrical standards, i.e. the noise voltage is traceable to the SI ampere, but any calibration drift in the sensing electronics used to measure this voltage is expected to have a negligible effect on the inferred temperature.

With an achievable uncertainty of about ±0.5°C (54) when the measurement is averaged over a few seconds, the most obvious application of practical Johnson noise thermometry is as a replacement for thermocouples. As the uncertainty can be decreased by measuring over a longer time period (it decreases with the square root of the time over which measurements are averaged) it may also have applications in temperature standards and calibration applications. The current prototype operates between room temperature and about 1000°C. Development is currently focusing on increasing the upper temperature by exploring new probe designs, and also different sensing materials which are sufficiently robust to withstand temperatures up to around 1200°C and whose resistance can be tuned. Experiments with a range of exotic sensing element materials are currently underway.

There are other types of temperature sensor which could, in time, be used for practical primary thermometry. Examples of these include Doppler broadening thermometry (58) which links thermal energy to the optical frequency of an atomic or molecular spectral line, and acoustic gas thermometry (59) which measures the speed of sound in a gas, and which was used to determine the Boltzmann constant for the redefinition of the kelvin with unprecedented accuracy (60). Practical implementations of these devices are in varying stages of development. The ultimate goal is to enable users to measure temperature directly, dispensing with the need for calibration and eliminating the problem of calibration drift.

Conclusion

The problem of calibration drift and its effects on process control, and the consequent impact on factors such as efficiency, product yield and safety have been described. Some new techniques to overcome it and retain traceability to national temperature standards have also been described, specifically self-validating thermocouples and embedded miniature reference cells for in situ traceable recalibration. In addition, a paradigm shift in the form of practical primary thermometry has been introduced by means of a practical Johnson noise thermometer, which enables direct measurement of temperature which does not rely on calibrations and does not exhibit calibration drift. Many of these developments are at the advanced prototype stage and are expected to enter the market in the next few years, offering the possibility of greatly improved process measurement and control.

References

  1. 1.
    G. Machin, Meas. Sci. Technol., 2018, 29, (2), 022001 LINK https://doi.org/10.1088/1361-6501/aa9ddb
  2. 2.
    G. Machin, Johnson Matthey Technol. Rev., 2023, 67, (1), 77 LINK https://technology.matthey.com/article/67/1/77-84/
  3. 3.
    H. Preston-Thomas, Metrologia, 1990, 27, (1), 3 LINK https://doi.org/10.1088/0026-1394/27/1/002
  4. 4.
    ‘General Requirements for the Competence of Testing and Calibration Laboratories’, ISO/IEC 17025:2017, International Organization for Standardization, Geneva, Switzerland, 2017 LINK https://www.iso.org/standard/66912.html
  5. 5.
    J. Pearce, Nat. Phys., 2017, 13, (1), 104 LINK https://doi.org/10.1038/nphys4005
  6. 6.
    M. Baker, Nature, 2016, 533, (7604), 452 LINK https://doi.org/10.1038/533452a
  7. 7.
    M. Sené, I. Gilmore and J.-T. Janssen, Nature, 2017, 547, (7664), 397 LINK https://doi.org/10.1038/547397a
  8. 8.
    G. Machin, AIP Conf. Proc., 2013, 1552, (1), 305 LINK https://doi.org/10.1063/1.4821383
  9. 9.
    E. R. Woolliams, G. Machin, D. H. Lowe and R. Winkler, Metrologia, 2006, 43, (6), R11 LINK https://doi.org/10.1088/0026-1394/43/6/r01
  10. 10.
    D. H. Lowe, A. D. W. Todd, R. Van den Bossche, P. Bloembergen, K. Anhalt, M. Ballico, F. Bourson, S. Briaudeau, J. Campos, M. G. Cox, D. del Campo, M. R. Dury, V. Gavrilov, I. Grigoryeva, M. L. Hernanz, F. Jahan, B. Khlevnoy, V. Khromchenko, X. Lu, G. Machin, J. M. Mantilla, M. J. Martin, H. C. McEvoy, B. Rougié, M. Sadli, S. G. R. Salim, N. Sasajima, D. R. Taubert, E. van der Ham, T. Wang, D. Wei, A. Whittam, B. Wilthan, D. J. Woods, J. T. Woodward, E. R. Woolliams, Y. Yamada, Y. Yamaguchi, H. W. Yoon and Z. Yuan, Metrologia, 2017, 54, (3), 390 LINK https://doi.org/10.1088/1681-7575/aa6eeb
  11. 11.
    F. Edler and A. C. Baratto, Metrologia, 2005, 42, (4), 201 LINK https://doi.org/10.1088/0026-1394/42/4/003
  12. 12.
    H. Ogura, M. Izuchi and M. Arai, Int. J. Thermophys., 2008, 29, (1), 210 LINK https://doi.org/10.1007/s10765-007-0320-x
  13. 13.
    R. Morice, P. Ridoux and J. R. Filtz, NCSL Int. Meas., 2008, 3, (1), 44 LINK https://doi.org/10.1080/19315775.2008.11721411
  14. 14.
    Y.-G. Kim, I. Yang, S. Y. Kwon and K. S. Gam, Metrologia, 2006, 43, (1), 67 LINK https://doi.org/10.1088/0026-1394/43/1/010
  15. 15.
    M. Tischler and M. J. Koremblit, ‘Miniature thermometric fixed points for thermocouple calibrations’, Temperature: Its Measurement and Control in Science and Industry, 5th Temperature Symposium, 15th–18th March, 1982, Washington DC, USA, AIP Conference Proceedings, AIP Publishing LLC, New York, USA, 1982, pp. 383–390
  16. 16.
    H. Ronsin, M. Elgourdo, G. Bonnier, M.V. Chattle, S.J. Read, G. Bongiovanni, R. Perissi, Ib. Wessel, M.J. de Groot and R. den Dekker, ‘Assessment of minicrucible fixed points for thermocouple calibration, through an international comparison’, Temperature: Its Measurement and Control in Science and Industry, 6th Temperature Symposium, 28th April–1st May, 1992, Toronto, Ontario, Canada, AIP Conference Proceedings, AIP Publishing LLC, New York, USA, 1992, pp. 1061–1068
  17. 17.
    S. Augustin, F. Bernhard, D. Boguhn, A. Donin, and H. Mammen, ‘Industrially Applicable Miniature Fixed Point Thermocouples’, 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, Berlin, Germany, 19th–21st June, 2001, TEMPMEKO 2001, ed. B. Fellmuth, Vol. 2, The Association of German Engineers (VDI), Dusseldorf, Germany and Association for Electrical, Electronic & Information Technologies eV (VDE), Frankfurt, Germany, 2002, pp. 3–8 LINK https://www.tib.eu/en/search/id/BLCP:CN044253136/industrially-applicable-miniature-fixed-point-thermocouples?cHash=d0cf42a538bc30a2afea878393c2c5b2
  18. 18.
    F. Bernhard, Proc. Estonian Acad. Sci. Eng., 2007, 13, (4), 320 LINK https://kirj.ee/public/Engineering/2007/issue_4/eng-2007-4-6.pdf
  19. 19.
    K. H. Kang, Y.-G. Kim, K. S. Gam and I. Yang, Meas. Sci. Technol., 2007, 18, (9), 3005 LINK https://doi.org/10.1088/0957-0233/18/9/035
  20. 20.
    H. Lehmann, Int. J. Thermophys., 2010, 31, (8–9), 1599 LINK https://doi.org/10.1007/s10765-010-0796-7
  21. 21.
    J. V. Pearce, O. Ongrai, G. Machin and S. J. Sweeney, Metrologia, 2010, 47, (1), L 1 LINK https://doi.org/10.1088/0026-1394/47/1/l01
  22. 22.
    J. V. Pearce, C. J. Elliott, G. Machin and O. Ongrai, AIP Conf. Proc., 2013, 1552, (1), 595 LINK https://doi.org/10.1063/1.4821396
  23. 23.
    ‘iTHERM TrustSens’, Endress + Hauser AG, Reinach, Switzerland: https://www.ch.endress.com/en/field-instruments-overview/temperature-measurement-thermometers-transmitters/TrustSens-infopage-en (Accessed on 21st October 2022)
  24. 24.
    D. Tucker, A. Andreu, C. Elliott, T. Ford, M. Neagu, G. Machin and J. Pearce, Meas. Sci. Technol., 2018, 29, (10), 105002 LINK https://doi.org/10.1088/1361-6501/aad8a8
  25. 25.
    C. J. Elliott, A. D. Greenen, D. Tucker, T. Ford and J. V Pearce, Int. J. Thermophys., 2017, 38, (9), 141 LINK https://doi.org/10.1007/s10765-017-2274-y
  26. 26.
    C. Tyson, C. McAndrew, P. L. Tuma, I. Pegg and A. Sarkar, Cytom. Part A, 2015, 87, (5), 393 LINK https://doi.org/10.1002/cyto.a.22631
  27. 27.
    J. P. G. van Brakel, ‘Robust Peak Detection Algorithm Using Z-Scores’, Version 2020-11-08, Stack Overflow Ltd, London, UK, 2014 LINK https://stackoverflow.com/questions/22583391/peak-signal-detection-in-realtime-timeseries-data/22640362#22640362
  28. 28.
    J. V. Pearce, R. I. Veltcheva, D. M. Peters, D. Smith and T. Nightingale, Meas. Sci. Technol., 2019, 30, (12), 124003 LINK https://doi.org/10.1088/1361-6501/ab38e4
  29. 29.
    D. Smith, D. Peters, T. Nightingale, J. Pearce and R. Veltcheva, Remote Sens., 2020, 12, (11), 1832 LINK https://doi.org/10.3390/rs12111832
  30. 30.
    V. I. Sapritsky, A. A. Burdakin, B. B. Khlevnoy, S. P. Morozova, S. A. Ogarev, A. S. Panfilov, V. N. Krutikov, G. E. Bingham, T. W. Humpherys, J. Tansock, A. V. Thurgood and V. Privalsky, J. Appl. Remote Sens., 2009, 3, (1), 033506 LINK https://doi.org/10.1117/1.3086288
  31. 31.
    J. Sun, X. Hao, F. Zeng, L. Zhang and X. Fang, Int. J. Thermophys., 2017, 38, (4), 47 LINK https://doi.org/10.1007/s10765-017-2181-2
  32. 32.
    X. P. Hao, J. P. Sun, C. Y. Xu, P. Wen, J. Song, M. Xu, L. Y. Gong, L. Ding and Z. L. Liu, Int. J. Thermophys., 2017, 38, (6), 90 LINK https://doi.org/10.1007/s10765-017-2223-9
  33. 33.
    A. Burdakin, B. Khlevnoy, M. Samoylov, V. Sapritsky, S. Ogarev, A. Panfilov, G. Bingham, V. Privalsky, J. Tansock and T. Humpherys, Metrologia, 2008, 45, (1), 75 LINK https://doi.org/10.1088/0026-1394/45/1/011
  34. 34.
    A. Burdakin, B. Khlevnoy, M. Samoylov, V. Sapritsky, S. Ogarev, A. Panfilov and S. Prokhorenko, Int. J. Thermophys., 2009, 30, (1), 20 LINK https://doi.org/10.1007/s10765-008-0441-x
  35. 35.
    P. J. Gero, J. A. Dykema and J. G. Anderson, J. Atmos. Ocean. Technol., 2008, 25, (11), 2046 LINK https://doi.org/10.1175/2008jtecha1100.1
  36. 36.
    T. S. Topham, H. Latvakoski and M. Watson, AIP Conf. Proc., 2013, 1552, (1), 993 LINK https://doi.org/10.1063/1.4819679
  37. 37.
    T. S. Topham, G. E. Bingham, H. Latvakoski, I. Podolski, V. S. Sychev and A. Burdakin, npj Microgravity, 2015, 1, 15009 LINK https://doi.org/10.1038/npjmgrav.2015.9
  38. 38.
    M. Fukasawa, H. Freeland, R. Perkin, T. Watanabe, H. Uchida and A. Nishina, Nature, 2004, 427, (6977), 825 LINK https://doi.org/10.1038/nature02337
  39. 39.
    B. Fellmuth, J. Fischer, G. Machin, S. Picard, P. P. M. Steur, O. Tamura, D. R. White and H. Yoon, Phil. Trans. R. Soc. A, 2016, 374, (2064), 20150037 LINK https://doi.org/10.1098/rsta.2015.0037
  40. 40.
    J. F. Qu, S. P. Benz, H. Rogalla, W. L. Tew, D. R. White and K. L. Zhou, Meas. Sci. Technol., 2019, 30, (11), 112001 LINK https://doi.org/10.1088/1361-6501/ab3526
  41. 41.
    P. Bramley, D. Cruickshank and J. Pearce, Int. J. Thermophys., 2017, 38, (2), LINK https://doi.org/10.1007/s10765-016-2156-8
  42. 42.
    J. F. Qu, S. P. Benz, H. Rogalla, W. L. Tew, D. R. White and K. L. Zhou, Meas. Sci. Technol., 2019, 30, (11), 112001 LINK https://doi.org/10.1088/1361-6501/ab3526
  43. 43.
    B. P. Kibble and G. H. Rayner, “Coaxial AC Bridges”, Adam Hilger Ltd, Bristol, UK, 1984
  44. 44.
    H. Brixy, R. Hecker, J. Jehmen, P. Barbonus and R. Hans, ‘Temperature Measurement’, Specialist Meeting on Gas-Cooled Reactor Core and High Temperature Instrumentation, Bowness-on-Windermere, UK, 15th–17th June, 1982, IAEA-TC-389/6-7, International Atomic Energy Commission, Vienna, Austria, 1982
  45. 45.
    H. Von Brixy and T. Kakuta, ‘Noise Thermometer’, JAERI Review 96-003, Japan Atomic Energy Research Institute, Ibaraki Prefecture, Japan, March, 1996, 83 pp LINK https://jopss.jaea.go.jp/pdfdata/JAERI-Review-96-003.pdf
  46. 46.
    R. J. Soulen, R. L. Rusby and D. Van Vechten, J. Low Temp. Phys., 1980, 40, (5–6), 553 LINK https://doi.org/10.1007/bf00119524
  47. 47.
    T. V. Blalock, M. J. Roberts and R. L. Shepard, ‘ In Situ Calibration of Nuclear Plant Resistance Thermometers Using Johnson Noise’, Industrial Temperature Measurement Symposium, Knoxville, USA, 10th–12th September 1984, CONF-8409109—7, DE85 004434, Oak Ridge National Laboratory, Oak Ridge, USA, 1985, 11 pp LINK https://www.osti.gov/servlets/purl/6061815
  48. 48.
    M. J. De Groot, J. F. Dubbeldam, G. S. Dhupia, M. V. Chattle, H. Brixy, and F. Edler, ‘Development of a High Temperature Resistance Thermometer using Noise Thermometry’, Proceedings of TEMPMEKO ’96, Torino, Italy, 10th–12th September, 1996, ed. P. Marcarino, Levrotto and Bella Libreria Editrice Universitaria, Torino, Italy, 1997, pp. 141–146
  49. 49.
    C. J. Borkowski and T. V Blalock, Rev. Sci. Instrum., 1974, 45, (2), 151 LINK https://doi.org/10.1063/1.1686578
  50. 50.
    H. Brixy, R. Hecker, J. Oehmen and E. Zimmermann, High Temp.-High Press., 1991, 23, (6), 625
  51. 51.
    K. Korsah, R. A. Kisner, C. L. Britton, P. Ramuhalli, D. W. Wootan, N. C. Anheier, A. A. Diaz, E. H. Hirt, R. B. Vlim, H. T. Chien, S. Bakhtiari, S. Sheen, S. Gopalsami, A. Heifetz, S. W. Tam, Y. Park, B. R. Upadhyaya and A. Stanford, “Assessment of Sensor Technologies for Advanced Reactors”, ORNL/TM-2016/337 R1, Oak Ridge National Laboratory, Oak Ridge, USA, August, 2016, 168 pp LINK https://doi.org/10.2172/1345781
  52. 52.
    D. Drung and C. Krause, Rev. Sci. Instrum., 2021, 92, (3), 034901 LINK https://doi.org/10.1063/5.0035673
  53. 53.
    ‘Johnson Noise Thermometer’, Metrosol Ltd, Paulerspury, UK: http://www.johnson-noise-thermometer.com (Accessed on 27th October 2022)
  54. 54.
    P. Bramley, D. Cruickshank and J. Aubrey, Meas. Sci. Technol., 2020, 31, (5), 054003 LINK https://doi.org/10.1088/1361-6501/ab58a6
  55. 55.
    ‘Electromagnetic Compatibility (EMC) – Part 4-3: Testing and Measurement Techniques – Radiated, Radio-Frequency, Electromagnetic Field Immunity Test’, IEC 61000-4-3:2020, International Electrotechnical Commission, Geneva, Switzerland, 2020 LINK https://webstore.iec.ch/publication/59849
  56. 56.
    S. O. Rice, Bell Syst. Tech. J., 1944, 23, (3), 282 LINK https://doi.org/10.1002/j.1538-7305.1944.tb00874.x
  57. 57.
    S. O. Rice, Bell Syst. Tech. J., 1945, 24, (1), 46 LINK https://doi.org/10.1002/j.1538-7305.1945.tb00453.x
  58. 58.
    L. Gianfrani, Philos. Trans. R. Soc. A, 2016, 374, (2064), 20150047 LINK https://doi.org/10.1098/rsta.2015.0047
  59. 59.
    M. R. Moldover, R. M. Gavioso, J. B. Mehl, L. Pitre, M. de Podesta and J. T. Zhang, Metrologia, 2014, 51, (1), R1 LINK https://doi.org/10.1088/0026-1394/51/1/r1
  60. 60.
    J. Fischer, B. Fellmuth, C. Gaiser, T. Zandt, L. Pitre, F. Sparasci, M. D. Plimmer, M. de Podesta, R. Underwood, G. Sutton, G. Machin, R. M. Gavioso, D. Madonna Ripa, P. P. M. Steur, J. Qu, X. J. Feng, J. Zhang, M. R. Moldover, S. P. Benz, D. R. White, L. Gianfrani, A. Castrillo, L. Moretti, B. Darquié, E. Moufarej, C. Daussy, S. Briaudeau, O. Kozlova, L. Risegari, J. J. Segovia, M. C. Martín and D. del Campo, Metrologia, 2018, 55, (2), R1 LINK https://doi.org/10.1088/1681-7575/aaa790

EMPIR Real-K project LINK https://real-k.aalto.fi/

Acknowledgements

We would like to thank Trevor Ford, Peter Cowley and Phill Williams of CCPI Europe Ltd, UK, for contributions on the self-validating thermocouples, Dan Peters and Dave Smith of RAL Space, UK, for contributions on the embedded phase-change cell, and Paul Bramley and David Cruickshank of Metrosol Ltd, UK, for contributions on Johnson noise thermometry. The miniature phase-change cell work was carried out under UK Space Agency Centre for Earth Observation Instrumentation contracts UKSA and CEOI-ST 11th Flagship project “Next Generation InfraRed calibration Sources (NGenIRS)” RP10G0435C204 and the NSTP3 Pathfinder project “Space Based Phase Change Cells for Calibration”, NSTP3-PF-044. LabVIEWTM is a trademark of National Instruments, UK.

The Authors

Jonathan Pearce is a Principal Research Scientist and head of contact thermometry in the Temperature & Humidity group at the National Physical Laboratory (NPL), UK, where he has been based for 16 years. He has authored 164 technical papers on applied physics and temperature metrology. He specialises in development of temperature standards for the realisation and dissemination of the SI unit of temperature, the kelvin, as well as provision of process monitoring and control solutions for industry, government and academia. He is the UK representative on the EURAMET Technical Committee for Thermometry (TC-T), serves on the Consultative Committee of Thermometry (CCT) working group for Contact Thermometry and coordinates preparation of CCT best practice guides and is a Fellow of the Institute of Physics.

Declan Tucker is a Higher Research Scientist in the Temperature & Humidity group at NPL. He is currently working with Jonathan Pearce, specialising in the design and characterisation of novel thermocouples, such as self-validating thermocouples, and has experience of installing and commissioning state-of-the-art thermocouple calibration facilities at other national measurement institutes. He is also involved in work characterising thermocouple inhomogeneity and is a member of the Institute of Physics.

Radka Veltcheva is a Senior Research Scientist with more than 20 years’ experience in contact thermometry at several European national measurement institutes, with 14 years at the UK’s National Physical Laboratory. She works on realisation, dissemination and traceability of the ITS-90, calibration of the temperature fixed point and standard thermometers, development of temperature standards for space applications, automation and control of scientific experiments and industrial measurement processes and LabVIEWTM software programming. She has published more than 30 papers on temperature metrology. She has participated in international key comparisons, and she is a member of two EURAMET Technical Committee for Thermometry (TC-T) working groups, on calibration and measurement capabilities and on digitalisation of calibrations. She is a member of the Institute for Measurement and Control.

Graham Machin is an NPL Senior Fellow with more than 30 years of thermometry experience. He has published more than 240 papers and has given numerous keynote addresses. He serves on the Consultative Committee of Thermometry (CCT), chairing the CCT working group for Non-contact Thermometry and the task group for Body Temperature Measurement. He chaired the EURAMET Technical Committee for Thermometry (2014–2018) and was president of the Institute of Measurement and Control (2018–2019). He was awarded the InstMC Callendar Medal for “outstanding contributions to temperature measurement” (2012) and the InstMC Sir Harold Hartley Medal for “outstanding contributions to the technology of measurement and control” (2021). He was elected Honorary Scientist, Chinese Academy of Sciences and Fellow of the Royal Academy of Engineering (2019).

Related articles

“Practical Guidelines for the Chemical Industry: Operation, Processes, and Sustainability in Modern Facilities”

Microbubble Intensification of Bioprocessing

Find an article

ArticleSearch