Mass Loss of Platinum-Rhodium Thermocouple Wires at 1324°C

Lower mass loss observed with enclosed wires


  • Sivahami Uthayakumaar, Stuart Davidson, Jonathan Pearce*
  • National Physical Laboratory, Hampton Road, Teddington, TW11 0LW, UK
  • *Email: jonathan.pearce@npl.co.uk

Article Synopsis

It is known that platinum-rhodium thermocouples exhibit mass loss when in the presence of oxygen at high temperatures due to the formation of volatile oxides of platinum and rhodium. The mass losses of platinum, Pt-6%Rh and Pt-30%Rh wires, commonly used for thermocouples, were considered in this paper to characterise the mass loss of wires of the three compositions due to formation and evaporation of the oxides PtO2 and RhO2 under the conditions that would be seen by thermocouples used at high temperature. For the tests, the wires were placed in thin alumina tubes to emulate the thermocouple format, and the measurements were performed in air at a temperature of 1324°C, i.e. with oxygen partial pressure of 21.3 kPa. It was found that the mass loss of the three wires increases linearly with elapsed time, consistent with other investigations, up to an elapsed time of about 150 h, but after that, a marked acceleration of the mass loss is observed. Remarkably, previous high precision studies have shown that a crossover after about 150 h at 1324°C is also observed in the thermoelectric drift of a wide range of platinum-rhodium thermocouples, and the current results are compared with those studies. The mass loss was greatest for Pt-30%Rh, followed by Pt6%Rh, then platinum.

Introduction

Platinum-rhodium thermocouples are used widely as temperature sensors (1) and are often used in industry where the accuracy and stability requirements are greater than those that can be provided by less expensive base metal thermocouples. Examples of such processes are iron and steel manufacturing (2), quartz glass manufacturing and aerospace heat treatment and casting. Platinum and rhodium are also important industrial catalysts, for example for the oxidation of ammonia to nitric oxide (3, 4).

At temperatures above about 1200°C, it is known that platinum-rhodium thermocouple wires form volatile oxides where the solid platinum or rhodium oxide films formed on the surface of the wires at lower temperatures evaporate (5). This causes the wires to lose mass (3, 68), which is the dominant cause of progressive instability and thermoelectric-inhomogeneity of the platinum-rhodium alloyed thermocouples. This instability arises because the vapour pressures and oxidation rates of platinum and rhodium are different. The overall behaviour also differs for each alloy (2, 6, 9). The departure of platinum and rhodium oxides from the wire at different rates causes a change in the composition of the thermoelements (4, 6) and hence calibration drift of the thermocouple.

An investigation conducted by Rubel et al. (4) studied the contribution of the volatile oxides to the mass loss of Pt-5%Rh and Pt-10%Rh alloy gauzes. The gauzes were exposed to oxygen over periods of 70 h to 300 h at 890°C and 1100°C. By using X-ray photoelectron spectroscopy (XPS), the type of volatile oxides produced were identified as platinum and rhodium oxides PtO2 and RhO2 respectively. It was also found that the rate of mass loss of rhodium oxide was greater with increasing rhodium content of the platinum-rhodium alloy.

Although the oxidation rates of platinum-rhodium alloys have not been investigated much in recent years, the effect of oxidation of the platinum group metals at high temperatures has been widely studied. Phillips, in particular, studied the mass loss of the platinum group metals osmium, palladium, ruthenium, platinum, iridium and rhodium, and evaluated the time dependence of the mass loss by considering metal powders placed in 1’’ billets over the temperature range from 600°C to 1400°C (10). He concluded that all the platinum group metals lost mass when oxidised in air and that the rate of mass loss was greater at higher temperatures. Notably, the rate of mass loss increased with temperature, and for a given temperature the mass loss was linear with elapsed time, provided the volatile oxides are formed from the corresponding metal.

Another key experiment to determine the mass loss of platinum metals was conducted by Crookes (11). The two temperatures considered in that investigation were 900°C and 1300°C. The mass loss was investigated by using crucibles made of platinum and rhodium, where the original mass of platinum and rhodium was about 150 g and 33 g respectively. The mass loss was measured at time intervals of 2 h for a total period of 30 h. The significance of the results from this investigation was that at 900°C the mass loss of the platinum and rhodium crucibles was found to be negligible. However, at 1300°C it was found that the mass loss of both platinum and rhodium increased linearly with time. This measurement showed that volatile oxides are formed and vaporise at high temperatures, and that these are formed above some threshold temperature (3, 11).

Jehn (5) also observed a linear dependence of the mass loss with elapsed time, though it was noted in that study that in other experiments extended over “extremely long time periods” a marked increase in the mass loss rate had been observed for platinum at 1300°C after about 60 h (12) and 400 h (13); no explanation was given.

One difficulty with previous measurements of the rate of mass loss of platinum and platinum-rhodium alloys is the very wide dispersion of results, which vary by several orders of magnitude. This may be due in part to the wide variation of geometries of the test conditions, and the wide variation of environments (such as ambient gas or vacuum). Furthermore, none are applicable to typical thermocouple formats, which comprise thin noble metal wires inserted in narrow ceramic (typically alumina) bores. In addition, no attempt has been made to link the rate of mass loss of the wires to thermocouple drift performance. The aim of the current investigation was to perform such mass loss measurements in a manner that is directly applicable to thermocouples, to better inform thermocouple drift models based on evaporation of platinum and rhodium oxides. Wires of Pt-30%Rh, Pt-6%Rh and platinum were investigated in ambient atmospheric conditions, within a typical alumina insulation tube used for thermocouples to make the test as realistic as possible in the context of thermocouple usage. A temperature of 1324°C was selected for the tests, because this is the melting temperature of the cobalt-carbon alloy, which is a well-established temperature reference point (a so-called ‘fixed point’) for calibrating thermocouples, and is used for in situ periodic calibration during long-term experimental thermoelectric drift tests.

In this article, the method is described, followed by a presentation of the results and an assessment of the associated uncertainties. The results are compared with thermocouple stability measurements where there are some intriguing qualitative similarities. Finally, some conclusions are drawn.

Method

Two identical wires were used for each type of wire to investigate the change of mass, where one wire was used as the control (kept at room temperature) and the other was the test wire exposed to high temperatures. This allows the change in mass to be calculated by comparing the masses of both wires. Measuring the mass change in this way reduces the influence of calibration drift and scale linearity of the mass balance. The wire diameter for all wires was 0.5 mm. The length of the platinum wire was 60 cm, while the Pt-6%Rh and Pt-30%Rh wires were 70 cm long.

Each of the wires was thoroughly cleaned using acetone and distilled water in order to remove any surface contamination prior to making mass measurements and heating. The wire masses were then measured using a Sartorius Supermicro S4 balance (Sartorius AG, Germany) with a resolution of 0.1 μg and a standard deviation of less than 1 μg. The measurement uncertainty of the mass comparison performed was approximately ±25 μg, largely due to random fluctuations in the readings. The balance was placed in a laboratory at a constant temperature (20°C ± 1°C), and the apparatus was kept in an isothermal enclosure to prevent the air from the surroundings from interfering with the measurements.

Prior measurements were made to determine how the surroundings affect the mass measurements using two scrap platinum wires. One wire was kept as a control, where the scales were tared, and the other was the test wire (which in subsequent measurements would be the one exposed to high temperatures). The air conditioning in the laboratory caused short-term temperature fluctuations and air currents which manifested as long stabilisation times and noise in the mass measurements, so the thermal isolation of the balance was progressively improved until the stabilisation time and random fluctuations were both minimised.

To facilitate measurement on the balance pan, the wire was coiled around a plastic former to minimise contamination of the surface; the former was removed before weighing. Mass measurements of the test wires were made by comparison with the corresponding control wire, which was used to tare the balance. The mass of the control and test wires was measured three times to determine the average mass and to account for, and characterise, the random fluctuations in the test readings.

For each type of test wire used, three single alumina tubes of length 40 cm and diameter 3 mm, with bore diameter approximately 1 mm, were used. The wire was threaded into the bore to emulate the thermocouple format. The assembly was then placed in a thermal annealing furnace, which was maintained at 1324°C with stability of approximately ±1°C. Before beginning the experiment, the alumina tubes were maintained at 1500°C in air for 6 h to drive off volatile impurities, in order to minimise contamination of the test wires.

The platinum and platinum-rhodium test wires were placed into the alumina tube bores by straightening the test wires (which are coiled for the mass measurements), while the control wires were kept coiled and stored in clean plastic bags.

The bores containing the test wires were placed in the furnace for designated intervals before being removed at the end of each interval and coiled again to perform the mass measurements.

Results

The uncertainty of the mass measurements was assessed by considering the individual contributions listed in Table I, which were summed in quadrature in a manner consistent with the International Organization for Standardization (ISO) Guide to the Expression of Uncertainty in Measurement (GUM) (14). The two statistical contributions, i.e. repeatability and reproducibility, were assessed by considering the standard deviation of repeated weighings in situ and of re-weighing the same item (i.e. removing from the balance and replacing) several times. The resolution, accuracy of the weights used to calibrate the balance, linearity and temperature of the balance reading were obtained from the manufacturer’s specifications. For general applicability of the results, the mass loss was expressed per unit surface area (mg cm–2); this was determined by dividing the mass loss values by the surface area of the wires. This yields an uncertainty (k = 2) of ±0.0027 mg cm–2 for the platinum wire and ±0.0023 mg cm–2 for the Pt-6%Rh and Pt-30%Rh wires.

Table I

Uncertainty Budget for Mass Measurementsa

Description Estimate, mg Probability distribution Divisor, mg Standard uncertainty, mg Comments
Repeatabilityb 0.0064 Normal 1.0000 0.0064 Highest recorded value
Reproducibilityc 0.0039 Normal 1.0000 0.0039 Highest recorded value
Resolution 0.0001 Rectangular 1.7321 0.0001 From specifications
Accuracy of calibration weights 0.0100 Normal 1.0000 0.0100 From specifications
Linearity 0.0003 Rectangular 1.7321 0.0002 From specifications
Temperature dependence 0.0002 Rectangular 1.7321 0.0001 From specificationsd
Combined uncertainty Normal 0.0125 Coverage factor k = 1 (67%)
Expanded uncertainty Normal 0.0250 Coverage factor k = 2 (95%)

a Standard uncertainty and expanded uncertainty correspond to normally distributed uncertainty with 67% (sometimes denoted one-sigma, or coverage factor k = 1) and 95% coverage probability (k = 2) respectively. Rectangular distributions correspond to the case where errors are equally likely across a range, such as resolution of the balance; the half-width of the distribution is divided by √3 as per the GUM guidance. Conversely, normal distributions generally correspond to cases where the errors are statistical in nature

b Repeated weighings in situ

c Re-weighing the same thing

d Maximum ambient temperature variation of ±1°C assumed

It was found that all the wires progressively lost mass during high temperature exposure. Figure 1 shows the mass loss per unit surface area as a function of time of exposure to high temperature. It can be seen that the mass loss is a linear function of elapsed time, at least up to about 150 h. Beyond that, however, a marked departure from linearity is seen, which suggests an acceleration of the mass loss per unit surface area. Intriguingly, previously reported very high precision drift tests of a set of different platinum-rhodium thermocouples (shown in Figure 2) show that they all exhibit a marked change in drift rate at about 150 h as well; beyond about 150 h the drift rate is constant for many hundreds of hours (15). The Pt-30%Rh appears to lose mass fastest, which is what might be expected since this wire contains more rhodium, and the vapour pressure of RhO2 is higher than that of PtO2 at this temperature (16).

Fig. 1.

Mass loss as a function of elapsed time at 1324°C for the three wires studied. Dashed line shows linear guide to the eye for data up to about 150 h; beyond that, there is a marked departure from linearity, where the mass loss appears to accelerate. The error bars (which are approximately the same size as the data points) correspond to the expanded uncertainty from Table I, converted to mass loss per unit area

Fig. 2.

Very high precision measurements of thermoelectric drift (in temperature terms) as a function of elapsed time at 1324°C for a variety of different thermocouples, after (14). Here the thermocouples are denoted by rhodium content. For instance, ‘6/0’ corresponds to Pt-6%Rh versus platinum. It can be seen that the drift is non-linear up to about 150 h, beyond which there is a ‘cross-over’ to a linear regime (shown by the dashed ‘guides to the eye’)

The fact that the mass change up to about 150 h is linear for the reported measurements is in qualitative agreement with previous results (5, 10, 11). However, the rate of mass loss of platinum, using the data of Jehn (5) and Phillips (10), was about 0.0049 mg cm–2 h–1 and 0.0048 mg cm–2 h–1 respectively at 1350°C. In the current study, the rate of mass loss up to 150 h was about 0.0006 mg cm–2 h–1, which is about an order of magnitude lower. This may be because in the present study the wire is enclosed in quite tight-fitting bores to emulate their use in thermocouples, so the reduced rate of mass loss may be due to the establishment of saturated vapour pressure above the wire (inhibiting further evaporation), or reduced oxygen present as the surface oxidation proceeds, or deposition of the oxide vapour or all of these effects. This is in contrast to the previous studies cited, where the samples were bulk (for example, crucibles, meshes, coupons) and the surface was exposed to the surroundings rather than enclosed in tight-fitting bores.

Discussion

The mass losses of platinum, Pt-6%Rh and Pt-30%Rh wires, commonly used for thermocouples, were considered as a function of elapsed time at 1324°C. The wires were placed in thin alumina tubes to emulate the thermocouple format. It was found that the mass loss of the three wires increases linearly with elapsed time, consistent with other investigations, up to an elapsed time of about 150 h, but after that, a marked acceleration of the mass loss is observed.

In the latter case, it is thought that, because the whole length of the thermocouple which develops an output is exposed to the full range of temperatures from room temperature to the measurement junction temperature, the behaviour in the first 150 h (17) arises from early development of microscopic ordering in the platinum-rhodium alloys (during exposure to temperatures below 600°C). This is followed by a slow coating of the platinum-rhodium leg by rhodium oxide (during exposure to temperatures between 600°C and 950°C) (1822). This is difficult to apply directly to the mass measurements because the all of the wire was held at 1324°C so the oxidation effects at lower temperatures do not occur, but it is conceivable that comparable ordering and coating behaviour is taking place during the slow increase and decrease of the setpoint temperature of the furnace. It is also possible that the mass loss simply varies non-linearly over the whole duration: the motivation for expressing the data in terms of a discontinuity at about 150 h duration has been guided by the previously observed marked increase in the rate of mass for platinum at 1300°C after about 60 h (12) and 400 h (13); and also by the rapid change in thermocouple drift rate at about 150 h. To draw a definitive conclusion on this point, more detailed measurements would be needed, perhaps with the wire exposed to the entire temperature range from ambient to 1324°C.

Conclusion

The rate of mass loss was approximately an order of magnitude less than that observed in previous studies, and the enclosed nature of the wires in the alumina insulation tubes is thought to be a cause of this. Previous high precision studies have shown that a change in the drift rate after about 150 h at 1324°C is also observed in the thermoelectric drift of a wide range of platinum-rhodium thermocouples, and the current results were compared with those studies. The mass loss was greatest for Pt-30%Rh, followed by Pt-6%Rh, then platinum.

References

  1. 1.
    J. V. Nicholas and D. R. White, “Traceable Temperatures: An Introduction to Temperature Measurement and Calibration”, John Wiley & Sons Ltd, Chichester, UK, 1994
  2. 2.
    B. Wu and G. Liu, Platinum Metals Rev., 1997, 41, (2), 81 LINK https://www.technology.matthey.com/article/41/2/81-85/
  3. 3.
    J. C. Chaston, Platinum Metals Rev., 1975, 19, (4), 135 LINK https://www.technology.matthey.com/article/19/4/135-140/
  4. 4.
    M. Rubel, M. Pszonicka and W. Palczewska, J. Mater. Sci., 1985, 20, (10), 3639 LINK https://doi.org/10.1007/bf01113771
  5. 5.
    H. Jehn, J. Less Common Met., 1984, 100, 321 LINK https://doi.org/10.1016/0022-5088(84)90072-9
  6. 6.
    F. Edler and P. Ederer, AIP Conf. Proc., 2013, 1552, (1), 532 LINK https://doi.org/10.1063/1.4819597
  7. 7.
    A. G. Knapton, Platinum Metals Rev., 1979, 23, (1), 2 LINK https://www.technology.matthey.com/article/23/1/2-13/
  8. 8.
    P. A. Kinzie, “Thermocouple Temperature Measurement”, John Wiley & Sons Inc, New York, USA, 1973, 278 pp
  9. 9.
    J. V. Pearce, Metrologia, 2020, 57, (2), 025009 LINK https://doi.org/10.1088/1681-7575/ab71b3
  10. 10.
    W. L. Phillips, Am. Soc. Met. Trans. Q., 1964, 57, (1), 33
  11. 11.
    W. Crookes, Proc. R. Soc. A: Math. Phys. Eng. Sci., 1912, 86, (589), 461 LINK https://doi.org/10.1098/rspa.1912.0040
  12. 12.
    E. V. Vasil’eva, M. I. Zakharova and E. M. Lazarev, Russ. J. Appl. Chem., 1974, 47, 1973
  13. 13.
    M. S. Tsirlip, B. I. Krasilov, V. B. Vladimirov and A. V. Byalobzheskin, Zashch. Met., 1973, 9, 478
  14. 14.
    Joint Committee for Guides in Metrology, ‘Evaluation of Measurement Data – Guide to the Expression of Uncertainty in Measurement’, JCGM 100:2008, International Organization for Standardization, Geneva, Switzerland, 2008 LINK https://www.iso.org/sites/JCGM/GUM/JCGM100/C045315e-html/C045315e.html?csnumber=50461
  15. 15.
    J. V. Pearce, A. D. Greenen, A. Smith and C. J. Elliott, Int. J. Thermophys., 2016, 38, (2), 26 LINK https://doi.org/10.1007/s10765-016-2159-5
  16. 16.
    C. B. Alcock and G. W. Hooper, Proc. R. Soc. A Math. Phys. Eng. Sci., 1960, 254, (1279), 551 LINK https://doi.org/10.1098/rspa.1960.0040
  17. 17.
    J. V. Pearce, D. Tucker, R. Rawal and L. Hutton, Meas. Sci. Technol., 2020, 31, (4), 044005 LINK https://doi.org/10.1088/1361-6501/ab48bd
  18. 18.
    E. S. Webster, Int. J. Thermophys., 2015, 36, (8), 1909 LINK https://doi.org/10.1007/s10765-015-1910-7
  19. 19.
    R. E. Bentley, Int. J. Thermophys., 1985, 6, (1), 83 LINK https://doi.org/10.1007/bf00505793
  20. 20.
    R. E. Bentley, “Theory and Practice of Thermoelectric Thermometry: Handbook of Temperature Measurement”, Vol. 3, Springer-Verlag Singapore Pte Ltd, Singapore, 1998
  21. 21.
    F. Jahan and M. Ballico, Int. J. Thermophys., 2007, 28, (6), 1832 LINK https://doi.org/10.1007/s10765-007-0304-x
  22. 22.
    F. Jahan and M. J. Ballico, Int. J. Thermophys., 2010, 31, (8–9), 1544 LINK https://doi.org/10.1007/s10765-010-0797-6

Acknowledgements

This work was performed under the auspices of the UK South East Physics Network (SEPnet) scheme, and as part of a European Metrology Programme for Innovation and Research (EMPIR) project to enhance process efficiency through improved temperature control, ‘17IND04 EMPRESS 2’. The EMPIR is jointly funded by the EMPIR participating countries within EURAMET and the European Union.

The Authors


Sivahami Uthayakumaar is currently a third-year PhD student at the University of York, UK, in the Nuclear Physics research group. She is specialising in nuclear structure with emphasis on isospin symmetry and the reasons that lead to this symmetry being broken. This research is studied in the proton-rich region of the chart of nuclides by analysing mirror nuclei (i.e. nuclei that have the proton and neutron numbers interchanged). The work reported in this paper was performed during a placement at the National Physical Laboratory (NPL), UK. She is a Member of the Institute of Physics.


Stuart Davidson is the Science Area Leader for Mechanical Metrology at the UK’s standards laboratory, the NPL. He has been responsible for some major advances in the field of mass measurement including the direct measurement of the density of air, weighing in vacuum and the use of surface analysis to predict the stability of mass standards. He is a past chair of the EURAMET Technical Committee for Mass and Related Quantities and current convenor of the Mass sub-committee and Strategy Working Group. He is also chair of the Consultative Committee for Mass and Related Quantities (CCM) Mass Working Group and the Task Group implementing the ‘new’ kilogram.


Jonathan Pearce is a principal research scientist and head of contact thermometry in the Temperature & Humidity group at the UK’s NPL, where he has been based for 15 years. He has authored 143 technical papers on applied physics and temperature metrology. He specialises in development of temperature standards for the realisation and dissemination of the SI unit of temperature, the kelvin, as well as provision of process monitoring and control solutions for industry, government and academia. He is the UK representative on the EURAMET Technical Committee for Thermometry (TC-T). He is a Fellow of the Institute of Physics.